Chroma Meter
Highly accurate, multi-functional, and user-friendly, Konica Minolta's
new Chroma Meters CR-400 and CR-410 provide support for measurement
using just the measuring head. |
|
Specifications
Principal Specifications CR-400 / 410
|
Name |
Chroma Meters Measuring Head |
|
Model |
CR-400 Head |
CR-410 Head |
|
Illuminating/viewing system |
d:0° (diffuse illumination/0° viewing angle; specular component
included)
(compliant with JIS Z 8722; including regular reflectance) |
Wide-area illumination/0° viewing angle; specular component
included) |
|
Detector |
Silicon photocells (6) |
|
Display range |
Y: 0.01% to 160.00% (reflectance) |
|
Light source |
Pulsed xenon lamp |
|
Measurement time |
1 sec. |
|
Minimum measurement interval |
3 sec. |
|
Battery performance |
Approx. 800 measurements (when using batteries under Konica
Minolta's testing conditions) |
|
Measurement/illumination area |
Ø 8 mm / Ø 11 mm |
Ø 50 mm/ Ø 53 mm |
|
Repeatability |
Within ΔE*ab 0.07 standard deviation (when the white calibration
plate is measured 30 times at intervals of 10 seconds) |
|
Inter-instrument agreement |
ΔE*ab: Within 0.6
Average of 12 BCRA Series II colors |
ΔE*ab: Within 0.8
Average of 12 BCRA Series II colors |
|
Observer condition |
CIE: 2° Standard Observer |
|
Illuminant condition |
CIE: C, D65 |
|
Tolerance judgment* |
Color difference tolerance (box tolerance and elliptical
tolerance) |
|
Color space/colorimetric data* |
XYZ, Yxy, L*a*b*, Hunter Lab, L*C*h, Munsell (Illuminant C only),
CMC (l:c), CIE1994, Lab99, LCh99, CIE2000, CIE WI/Tw (Illuminant
D65 only), WI ASTM E313 (Illuminant C only), YI ASTM D1925
(Illuminant C only), YI ASTM E313 (Illuminant C only), User index
(Up to 6 can be registered from computer) |
|
Languages* (Display) |
LCD: English (default), German, French, Italian, Spanish, Japanese |
|
Storable data sets |
1,000 (Measuring head and data processor save different data) |
|
Color-difference target colors |
100 |
|
Calibration channels* |
20 channels (ch00: White calibration; ch01 to ch19: User
calibration ) |
|
Display* |
Chroma values; color difference values; PASS/WARN/FAIL display |
|
Interface / Baud rate* |
RS-232 compliant (for data processor/PC)
* Baud rate: 4,800, 9,600, 19,200 (bps); Set at 9600 when shipped
from factory |
|
Power source |
Four AAA size alkaline or Ni-MH batteries
AC Adapter AC-A305 AC100 - 240 V 50/60Hz |
|
Dimensions (W × H × D) |
102 × 244 × 63 mm |
|
Weight |
Approximately 550 g
(including 4 AAA size batteries and not including RS-232C cable) |
Approximately 570 g
(including 4 AAA size batteries and not including RS-232C cable) |
|
Operating temperature/humidity range |
0 - 40°C, relative humidity 85% or less with no condensation |
|
Storage temperature/humidity range |
-20 - 40°C, relative humidity 85% or less with no condensation |
|
Other |
LCD backlight ON/OFF function (When on, backlight stays ON for 30
seconds after last key or measurement operation) |
|
Standard accessories |
White Calibration Plate CR-A43; Protective Cap CR-A72; Wrist Strap
CR-A73; AC Adapter AC-A305; 4 AAA size batteries |
White Calibration Plate CR-A44; Protective Cap CR-A72; Wrist Strap
CR-A73; AC Adapter AC-A305; 4 AAA size batteries |
|
Optional accessories |
Hard Case CR-A103; RS-232C Cable CR-A102 (for PC); Granular
Materials Attachment CR-A50; CR-400 Utility Software CR-S4w; Color
Data Software SpectraMagic NX CM-S100w ; Glass Light-Projection
Tube CR-A33a/f; Light-Projection Tube CR-A33c/d; Pivoting Base
CR-A12 |
Hard Case CR-A103; RS-232C Cable CR-A102 (for PC); Granular
Materials Attachment CR-A50; CR-400 Utility Software CR-S4w; Color
Data Software SpectraMagic NX CM-S100w ; Glass Light-Projection
Tube CR-A33e |
The "*" mark indicates functions available only when the Data
Processor is connected or when the required settings have been made.
|
Name |
Data Processor |
|
Model |
DP-400 |
|
Display range |
Y: 0.01% to 160.00% (reflectance) |
|
Measurement time* |
1 sec. |
|
Minimum measurement interval* |
3 sec. |
|
Battery performance |
Approx. 800 measurements (when using batteries under Konica
Minolta's testing conditions) |
|
Illuminant |
C, D65 |
|
Display |
Chroma values; color difference values; color difference graphs;
PASS/WARN/FAIL display |
|
Tolerance judgment* |
Color-difference tolerance (box tolerance and elliptical
tolerance) Only for display function |
|
Color space/colorimetric data* |
XYZ, Yxy, L*a*b*, Hunter Lab, L*C*h, Munsell (Illuminant C only),
CMC (l:c), CIE1994, Lab99, LCh99, CIE2000, CIE WI/Tw (Illuminant
D65 only), WI ASTM E313 (Illuminant C only), YI ASTM D1925
(Illuminant C only), YI ASTM E313 (Illuminant C only), User index
(Up to 6 registered in the measuring head can be used) |
|
Languages* (Display) |
LCD: English (default), German, French, Italian, Spanish, Japanese |
|
Storable data sets |
Max. 2000 pieces of data (divided into 100 pages)
Deletion and Undoing of selected stored data (one or all data) are
possible |
|
Color-difference target colors* |
Only for the operating function (100 pieces of data when the
measuring head is connected; input of measurement values or
numeric) (Independent from page function) |
|
Calibration channels* |
Only for the operating function (20 channels when the measuring
head is connected)
(Channel 00: White calibration; Channels 01 through 19: User
calibration) |
|
Page function |
100 pages |
|
Display |
Dot-matrix LCD with backlight (16 characters x 9 lines + 1 line
for icon display) Contrast adjustment |
|
Printer |
384-dot line thermal printer (can also print graphs); Automatic
printing for each measurement (can be set not to print) |
|
Statistical function |
Maximum, minimum, average, and standard deviation |
|
Automatic measurement* |
Date and time display: Year, month, day, hour, minute
Timer: 3 seconds to 99 minutes (Some measurement modes require
more than 3 seconds) |
|
Interface |
RS-232C compliant Baud rate (bps): 19200 fixed (when connected to
PC); when measuring head is connected, baud rate is automatically
set to that of the measuring head. |
|
Power source |
Four AA-size alkaline or Ni-MH batteries
AC Adapter AC-A305 AC100 - 240 V 50/60Hz |
|
Dimensions (W × H × D) |
102 × 244 × 63 mm |
|
Weight |
Approx. 600 g (not including batteries and paper) |
|
Operating temperature/humidity range |
0 - 40°C, relative humidity 85% or less with no condensation |
|
Storage temperature/humidity range |
-20 - 40°C, relative humidity 85% or less with no condensation |
|
Other specifications* |
User calibration function (multi-calibration/manual calibration);
Measurements for automatic averaging function; Print on/off
function; CR-400 measurement data import function; All color space
print on/off function; Data protection on/off function; Backlight
on/off function; Buzzer on/off function; Display color limit
function; Remote mode (stored data output); Character input
function (alphanumeric) |
|
Standard accessories |
Roll paper (1); RS-232C Cable CR-A101 (Head-DP) |
|
Optional accessories |
Shoulder Strap SS-01; Hard Case CR-A103; Roll Paper DP-A22 (5
rolls); RS-232C Cable CR-A102 (for PC); CR-400 Utility Software
CR-S4w; Color Data Software SpectraMagic NX CM-S100w |
The "*" mark indicates functions that can be used only when the
measuring head is connected.
Specifications are subject to change without prior notice.
|
|